The Patch-Clamp Field Guide
A field guide for doctoral researchers

From 5 MΩ to 1 GΩ and through the membrane

Patch clamp holds a membrane at a voltage you choose and reports the current it takes to keep it there, down to single channels passing a few picoamps. This guide pairs the physics with simulators you can drive, so you can see what the numbers on your amplifier mean before they go into a figure.

12 sections8 simulators25+ equationsSI & bench units
§1 · Seal test · +10 mV pulsePipette in bath
R (pulse)
5.0 MΩ
I steady
2.00 nA
C transient
none

Open tip in the bath. Ohm’s law gives the pipette resistance directly: 10 mV / 2 nA = 5 MΩ.

§2 · Recording configurations

Five ways to hold a patch

Every configuration starts from the cell-attached gigaseal. What you do next decides which face of the membrane touches which solution, and what you can measure.

Geometry
extracellular-likeintracellular-likerecorded membrane

§3 · Whole-cell equivalent circuit

Three components explain most of your membrane test

After break-in, the amplifier sees an access (series) resistance Rs in series with the cell, which is a membrane capacitance Cm in parallel with a membrane resistance Rm.

amp Rs Cm Rm Vm bath ground

Step the command by ΔV. At the first instant the capacitor is a short, so only Rs limits the current. As Cm charges, the current relaxes to the steady value set by Rs + Rm.

Current after a voltage step
\[ I(t)=\frac{\Delta V}{R_s+R_m}+\left(\frac{\Delta V}{R_s}-\frac{\Delta V}{R_s+R_m}\right)e^{-t/\tau},\qquad \tau=\frac{R_sR_m}{R_s+R_m}\,C_m \]
What your acquisition software inverts
\[ R_s=\frac{\Delta V}{I_{\text{peak}}},\qquad R_{\text{in}}=\frac{\Delta V}{I_{\text{ss}}}=R_s+R_m,\qquad C_m=\tau\left(\frac{1}{R_s}+\frac{1}{R_m}\right)\]

Why Rs is the enemy

The membrane only sees the command voltage minus the drop across Rs. The voltage error scales with the current you are trying to measure, and RsCm low-pass filters the clamp itself.

\[ V_m = V_{\text{cmd}} - I\,R_s(1-\alpha),\qquad \tau_{\text{clamp}}=R_s(1-\alpha)\,C_m,\qquad f_{-3\text{dB}}=\frac{1}{2\pi R_s(1-\alpha)C_m}\]

α = fractional series-resistance compensation (typically 0.6–0.85 with prediction; more tends to ring).

Unit trick worth memorising: MΩ × pF = µs, mV / MΩ = nA, nA × MΩ = mV.

§3 · Membrane test simulatorΔ10 mV
ideal (unfiltered)recorded after filter
I peak (rec.)
I steady
τ membrane
Rs estimated
Cm estimated
V error
clamp f−3dB

§4 · Electrochemistry

Where the current wants to go

Your pipette solution replaces the cytoplasm within minutes of break-in, so the reversal potentials in your experiment are the ones you mixed, not the cell’s own.

Nernst equation (one ion)
\[ E_X=\frac{RT}{z_XF}\ln\frac{[X]_o}{[X]_i}\;\approx\;\frac{61.5\ \text{mV}}{z_X}\log_{10}\frac{[X]_o}{[X]_i}\ \ (37\,^{\circ}\mathrm{C}) \]
Goldman–Hodgkin–Katz voltage equation
\[ V_{\text{rev}}=\frac{RT}{F}\ln\frac{P_K[K]_o+P_{Na}[Na]_o+P_{Cl}[Cl]_i}{P_K[K]_i+P_{Na}[Na]_i+P_{Cl}[Cl]_o} \]
Ohmic driving force
\[ I_X = g_X\,(V_m-E_X) \]

Liquid junction potential

Where pipette solution meets bath, ions of unequal mobility set up a junction potential. It is nulled when you zero the offset in the bath, then vanishes at seal formation, leaving a systematic error. With K-gluconate internal against ACSF it is typically around +13 to +16 mV, so the true potential is more negative than the recorded one.

\[ V_m = V_{\text{recorded}} - V_{LJ} \]

Sign convention of Neher (1992); calculate VLJ for your own solutions (e.g. with the generalised Henderson equation) and report whether you corrected.

§4 · Reversal potential calculator
Ion[out] mM[in] mMP / PKEion
K⁺1
Na⁺
Cl⁻
RT/F
VGHK (resting)
Cl⁻ current at −70 mV

Concentrations stand in for activities; activity coefficients near 0.75 at physiological ionic strength mostly cancel in the ratio.

§5 · Voltage-dependent gating

Reading an I–V curve

Voltage-clamp families are summarised as peak current against step voltage. Dividing out the driving force gives conductance, which you fit with a Boltzmann function.

Boltzmann activation
\[ \frac{G}{G_{\max}}=\frac{1}{1+\exp\!\left(\dfrac{V_{1/2}-V}{k}\right)},\qquad k=\frac{RT}{z_gF} \]
Current–voltage relation
\[ I(V)=G_{\max}\,\frac{G}{G_{\max}}(V)\,\left(V-E_{\text{rev}}\right) \]
Hodgkin–Huxley formulation
\[ C_m\frac{dV}{dt}=-\bar g_{Na}m^3h\,(V-E_{Na})-\bar g_Kn^4\,(V-E_K)-g_L(V-E_L)+I_{\text{inj}} \] \[ \frac{dx}{dt}=\alpha_x(V)(1-x)-\beta_x(V)\,x,\qquad x_\infty=\frac{\alpha_x}{\alpha_x+\beta_x},\ \ \tau_x=\frac{1}{\alpha_x+\beta_x} \]

The slope factor k reports the effective gating charge: at 22 °C, k ≈ 25.4 mV / zg. A k of 6 mV means roughly four elementary charges move across the field.

Before fitting: subtract leak and capacitive currents (P/N protocol), and check that the peak current times residual Rs stays below ~5 mV, or the apparent V1/2 shifts and the curve steepens.

§5 · I–V explorer
§6 · Current clamp

Injecting current, reading voltage

In current clamp the cell charges like an RC circuit. The same Rs that troubles voltage clamp now adds an instantaneous voltage jump, which bridge balance subtracts.

Passive membrane equation
\[ C_m\frac{dV}{dt}=-\frac{V-V_{\text{rest}}}{R_m}+I_{\text{inj}}\;\;\Rightarrow\;\; \Delta V(t)=I_{\text{inj}}R_m\left(1-e^{-t/\tau_m}\right),\ \ \tau_m=R_mC_m \]
What the amplifier reports
\[ V_{\text{rec}}=V_m+I_{\text{inj}}\,R_s\,(1-\beta) \]

β = bridge balance fraction. Set it on a small hyperpolarising step until the instantaneous jump disappears.

Specific membrane properties
\[ C_m = c_m A,\quad c_m\approx 1\ \mu\text{F}\,\text{cm}^{-2}\;\Rightarrow\;\sim 10\ \text{pF per }1000\ \mu\text{m}^2,\qquad \tau_m = r_m c_m \]

A single exponential only fits a compact cell. Neurons with dendrites show a fast component from charge redistributing along the cable, so fit two exponentials and use the slower one as τm.

§6 · Current-step simulatorVrest −70 mV
true Vmrecorded
τm
ΔV steady
bridge artefact
§7 · Single channels & fluctuation analysis

Counting channels you cannot see

Macroscopic current is the sum of stochastic openings. Its mean and variance together reveal the single-channel current i and the number of channels N.

Macroscopic current
\[ \langle I\rangle = N\,i\,P_o \]
Binomial variance (Sigworth 1980)
\[ \sigma_I^2 = N i^2 P_o(1-P_o) = i\,\langle I\rangle-\frac{\langle I\rangle^2}{N} \]
Unitary conductance and dwell times
\[ \gamma = \frac{i}{V-E_{\text{rev}}},\qquad f(t)=\frac{1}{\tau}e^{-t/\tau},\qquad \tau_{\text{open}}=\frac{1}{\sum k_{\text{exit}}} \]

Non-stationary noise analysis repeats the same step many times, computes ensemble mean and variance at each time point, and fits the parabola. The initial slope gives i; the curvature gives N. Maximum variance falls at Po = 0.5.

If your data never climb past the apex, Po,max is below 0.5 and N is poorly constrained. Report confidence intervals.

§7 · Mean–variance parabolaγ at −60 mV:
⟨I⟩
σ²
σ² / ⟨I⟩
§8 · Noise, filtering, digitisation

Why the seal must be a gigaseal

The seal resistance is a thermal noise source in parallel with your patch. Its current noise falls as the square root of resistance, which is why single-channel recording only became possible with gigaohm seals.

Johnson (thermal) current noise
\[ i_{\text{rms}}=\sqrt{\frac{4k_BTB}{R}} \]
Filter and sampling
\[ t_{10\text{–}90}\approx\frac{0.3321}{f_c}\ \ (\text{Bessel / Gaussian}),\qquad f_s\ge 2f_c\ \ (\text{Nyquist}),\ \text{in practice}\ f_s\approx 5\text{–}10\,f_c \]
Signal-to-noise for a single channel
\[ \text{SNR}=\frac{i}{i_{\text{rms}}}\ \gtrsim\ 5\ \text{for reliable event detection} \]

Other sources add in quadrature: amplifier headstage noise, pipette capacitance acting with the pipette’s thermal voltage noise (reduced by coating with Sylgard and keeping the bath level low), and 50 Hz mains pickup.

§8 · Seal noise calculatorT 22 °C
irms seal
rise time
SNR
fs / fc
§9 · At the rig

From pulled glass to a stable recording

The order matters. Typical values are for whole-cell recording from neurons in acute slices; adjust for your preparation.

  1. Step 1

    Pull & fill

    Borosilicate glass, fire-polished for single-channel work. Filter internal (0.2 µm), fill tip then back-fill, remove bubbles.

    Rp 3–7 MΩ whole-cell · 5–15 MΩ single-channel
  2. Step 2

    Positive pressure, enter bath

    Keep the tip clean. Zero the pipette offset and read Rp from the test pulse.

    offset → 0 mV
  3. Step 3

    Approach

    Descend under visual control until a dimple forms on the membrane and the test-pulse resistance rises.

    R up by ~10–50 %
  4. Step 4

    Seal

    Release pressure, apply gentle suction, step holding toward −60 to −70 mV to help sealing.

    Rseal ≥ 1 GΩ
  5. Step 5

    Cancel fast capacitance

    Null pipette capacitance transients (Cfast) while still cell-attached.

    Cp ~ 3–8 pF
  6. Step 6

    Break in

    Brief suction pulses or a zap. Large capacitive transients appear and holding current shifts.

    watch for a sudden seal loss
  7. Step 7

    Compensate & record parameters

    Cancel Cslow, read Rs, Rm, Cm. Apply Rs compensation with prediction in voltage clamp; bridge balance in current clamp.

    α 60–85 %
  8. Step 8

    Monitor throughout

    Repeat the membrane test between protocols. Log Rs, holding current and Vrest over time and set exclusion rules before you look at the results.

    ΔRs < 20 %
§10 · Quality control & troubleshooting

Decide your inclusion criteria first

Common thresholds from the whole-cell literature. They depend on cell type, so justify yours in the methods and apply them blind to condition.

ParameterGoodBorderlineExcludeWhy it matters
Seal before break-in≥ 1 GΩ0.5–1 GΩ< 0.5 GΩLeak shunts Rm and depolarises the cell
Access resistance Rs< 15 MΩ15–25 MΩ> 25 MΩVoltage error I·Rs and slow clamp
Rs drift during recording< 10 %10–20 %> 20 %Changes amplitude and kinetics over time
Vrest (neurons)≤ −60 mV−50 to −60 mV> −50 mVIndicates cell health and seal quality
Holding current at −70 mV> −100 pA−100 to −200 pA< −200 pALarge inward leak signals a damaged cell
Peak I × residual Rs< 5 mV5–10 mV> 10 mVDistorts I–V and activation curves
The seal will not form
  • Dirty tip: raise positive pressure, change pipette after every attempt.
  • Unhealthy cells or debris on the surface: clean the cell with a stream from the pipette before approach.
  • Osmolarity mismatch: keep internal ~10–20 mOsm below external.
  • Rough tip: fire-polish; check for broken tips under high magnification.
The cell is lost at break-in
  • Suction too strong or too long: use short, graded pulses or the amplifier zap.
  • Holding too depolarised: break in at −60 to −70 mV.
  • Tip too large for the cell: raise Rp.
Rs rises during the recording
  • Membrane resealing: small, gentle suction can reopen access.
  • Drift of the pipette or slice: check the manipulator and perfusion flow.
  • Exclude the cell if the change passes your threshold; do not correct post hoc without saying so.
Ringing or oscillation when compensating
  • Too much Rs compensation or prediction: back off by 5–10 %.
  • Re-null Cfast and Cslow first; lower bath level to cut pipette capacitance.
  • Increase the compensation lag (e.g. 10–20 µs).
50 Hz hum and drifting baseline
  • Star grounding to a single point; check the Faraday cage and remove unused mains devices.
  • Rechlorinate Ag/AgCl wires and the bath electrode; use an agar bridge if the bath Cl⁻ changes.
  • Keep the perfusion line away from the headstage and avoid dripping.
§11 · Reference

Equation sheet

Every relation in this guide, with a button that copies the LaTeX source for your thesis or slides.

Constants: R = 8.314 J mol⁻¹ K⁻¹, F = 96 485 C mol⁻¹, kB = 1.381 × 10⁻²³ J K⁻¹, e = 1.602 × 10⁻¹⁹ C. RT/F = 25.4 mV at 22 °C, 26.7 mV at 37 °C.

§12 · Primary sources & further reading

Where these equations come from

  1. Neher E, Sakmann B (1976). Single-channel currents recorded from membrane of denervated frog muscle fibres. Nature 260:799–802. doi:10.1038/260799a0
  2. Hamill OP, Marty A, Neher E, Sakmann B, Sigworth FJ (1981). Improved patch-clamp techniques for high-resolution current recording from cells and cell-free membrane patches. Pflügers Arch 391:85–100. doi:10.1007/BF00656997
  3. Hodgkin AL, Huxley AF (1952). A quantitative description of membrane current and its application to conduction and excitation in nerve. J Physiol 117:500–544. doi:10.1113/jphysiol.1952.sp004764
  4. Goldman DE (1943). Potential, impedance, and rectification in membranes. J Gen Physiol 27:37–60. doi:10.1085/jgp.27.1.37
  5. Hodgkin AL, Katz B (1949). The effect of sodium ions on the electrical activity of the giant axon of the squid. J Physiol 108:37–77. doi:10.1113/jphysiol.1949.sp004310
  6. Sigworth FJ (1980). The variance of sodium current fluctuations at the node of Ranvier. J Physiol 307:97–129.
  7. Neher E (1992). Correction for liquid junction potentials in patch clamp experiments. Methods Enzymol 207:123–131.
  8. Barry PH, Lynch JW (1991). Liquid junction potentials and small cell effects in patch-clamp analysis. J Membr Biol 121:101–117.
  9. Armstrong CM, Gilly WF (1992). Access resistance and space clamp problems associated with whole-cell patch clamping. Methods Enzymol 207:100–122.
  10. Sakmann B, Neher E, eds. (1995). Single-Channel Recording, 2nd ed. Springer. The standard reference for noise, filtering and dwell-time analysis.
  11. Molecular Devices. The Axon Guide: Electrophysiology and Biophysics Laboratory Techniques. Free amplifier-level treatment of compensation, noise and grounding.