From 5 MΩ to 1 GΩ and through the membrane
Patch clamp holds a membrane at a voltage you choose and reports the current it takes to keep it there, down to single channels passing a few picoamps. This guide pairs the physics with simulators you can drive, so you can see what the numbers on your amplifier mean before they go into a figure.
Open tip in the bath. Ohm’s law gives the pipette resistance directly: 10 mV / 2 nA = 5 MΩ.
Five ways to hold a patch
Every configuration starts from the cell-attached gigaseal. What you do next decides which face of the membrane touches which solution, and what you can measure.
Three components explain most of your membrane test
After break-in, the amplifier sees an access (series) resistance Rs in series with the cell, which is a membrane capacitance Cm in parallel with a membrane resistance Rm.
Step the command by ΔV. At the first instant the capacitor is a short, so only Rs limits the current. As Cm charges, the current relaxes to the steady value set by Rs + Rm.
Why Rs is the enemy
The membrane only sees the command voltage minus the drop across Rs. The voltage error scales with the current you are trying to measure, and RsCm low-pass filters the clamp itself.
α = fractional series-resistance compensation (typically 0.6–0.85 with prediction; more tends to ring).
Unit trick worth memorising: MΩ × pF = µs, mV / MΩ = nA, nA × MΩ = mV.
Where the current wants to go
Your pipette solution replaces the cytoplasm within minutes of break-in, so the reversal potentials in your experiment are the ones you mixed, not the cell’s own.
Liquid junction potential
Where pipette solution meets bath, ions of unequal mobility set up a junction potential. It is nulled when you zero the offset in the bath, then vanishes at seal formation, leaving a systematic error. With K-gluconate internal against ACSF it is typically around +13 to +16 mV, so the true potential is more negative than the recorded one.
Sign convention of Neher (1992); calculate VLJ for your own solutions (e.g. with the generalised Henderson equation) and report whether you corrected.
| Ion | [out] mM | [in] mM | P / PK | Eion |
|---|---|---|---|---|
| K⁺ | 1 | |||
| Na⁺ | ||||
| Cl⁻ |
Concentrations stand in for activities; activity coefficients near 0.75 at physiological ionic strength mostly cancel in the ratio.
Reading an I–V curve
Voltage-clamp families are summarised as peak current against step voltage. Dividing out the driving force gives conductance, which you fit with a Boltzmann function.
The slope factor k reports the effective gating charge: at 22 °C, k ≈ 25.4 mV / zg. A k of 6 mV means roughly four elementary charges move across the field.
Before fitting: subtract leak and capacitive currents (P/N protocol), and check that the peak current times residual Rs stays below ~5 mV, or the apparent V1/2 shifts and the curve steepens.
Injecting current, reading voltage
In current clamp the cell charges like an RC circuit. The same Rs that troubles voltage clamp now adds an instantaneous voltage jump, which bridge balance subtracts.
β = bridge balance fraction. Set it on a small hyperpolarising step until the instantaneous jump disappears.
A single exponential only fits a compact cell. Neurons with dendrites show a fast component from charge redistributing along the cable, so fit two exponentials and use the slower one as τm.
Counting channels you cannot see
Macroscopic current is the sum of stochastic openings. Its mean and variance together reveal the single-channel current i and the number of channels N.
Non-stationary noise analysis repeats the same step many times, computes ensemble mean and variance at each time point, and fits the parabola. The initial slope gives i; the curvature gives N. Maximum variance falls at Po = 0.5.
If your data never climb past the apex, Po,max is below 0.5 and N is poorly constrained. Report confidence intervals.
Why the seal must be a gigaseal
The seal resistance is a thermal noise source in parallel with your patch. Its current noise falls as the square root of resistance, which is why single-channel recording only became possible with gigaohm seals.
Other sources add in quadrature: amplifier headstage noise, pipette capacitance acting with the pipette’s thermal voltage noise (reduced by coating with Sylgard and keeping the bath level low), and 50 Hz mains pickup.
From pulled glass to a stable recording
The order matters. Typical values are for whole-cell recording from neurons in acute slices; adjust for your preparation.
- Step 1
Pull & fill
Borosilicate glass, fire-polished for single-channel work. Filter internal (0.2 µm), fill tip then back-fill, remove bubbles.
Rp 3–7 MΩ whole-cell · 5–15 MΩ single-channel - Step 2
Positive pressure, enter bath
Keep the tip clean. Zero the pipette offset and read Rp from the test pulse.
offset → 0 mV - Step 3
Approach
Descend under visual control until a dimple forms on the membrane and the test-pulse resistance rises.
R up by ~10–50 % - Step 4
Seal
Release pressure, apply gentle suction, step holding toward −60 to −70 mV to help sealing.
Rseal ≥ 1 GΩ - Step 5
Cancel fast capacitance
Null pipette capacitance transients (Cfast) while still cell-attached.
Cp ~ 3–8 pF - Step 6
Break in
Brief suction pulses or a zap. Large capacitive transients appear and holding current shifts.
watch for a sudden seal loss - Step 7
Compensate & record parameters
Cancel Cslow, read Rs, Rm, Cm. Apply Rs compensation with prediction in voltage clamp; bridge balance in current clamp.
α 60–85 % - Step 8
Monitor throughout
Repeat the membrane test between protocols. Log Rs, holding current and Vrest over time and set exclusion rules before you look at the results.
ΔRs < 20 %
Decide your inclusion criteria first
Common thresholds from the whole-cell literature. They depend on cell type, so justify yours in the methods and apply them blind to condition.
| Parameter | Good | Borderline | Exclude | Why it matters |
|---|---|---|---|---|
| Seal before break-in | ≥ 1 GΩ | 0.5–1 GΩ | < 0.5 GΩ | Leak shunts Rm and depolarises the cell |
| Access resistance Rs | < 15 MΩ | 15–25 MΩ | > 25 MΩ | Voltage error I·Rs and slow clamp |
| Rs drift during recording | < 10 % | 10–20 % | > 20 % | Changes amplitude and kinetics over time |
| Vrest (neurons) | ≤ −60 mV | −50 to −60 mV | > −50 mV | Indicates cell health and seal quality |
| Holding current at −70 mV | > −100 pA | −100 to −200 pA | < −200 pA | Large inward leak signals a damaged cell |
| Peak I × residual Rs | < 5 mV | 5–10 mV | > 10 mV | Distorts I–V and activation curves |
The seal will not form
- Dirty tip: raise positive pressure, change pipette after every attempt.
- Unhealthy cells or debris on the surface: clean the cell with a stream from the pipette before approach.
- Osmolarity mismatch: keep internal ~10–20 mOsm below external.
- Rough tip: fire-polish; check for broken tips under high magnification.
The cell is lost at break-in
- Suction too strong or too long: use short, graded pulses or the amplifier zap.
- Holding too depolarised: break in at −60 to −70 mV.
- Tip too large for the cell: raise Rp.
Rs rises during the recording
- Membrane resealing: small, gentle suction can reopen access.
- Drift of the pipette or slice: check the manipulator and perfusion flow.
- Exclude the cell if the change passes your threshold; do not correct post hoc without saying so.
Ringing or oscillation when compensating
- Too much Rs compensation or prediction: back off by 5–10 %.
- Re-null Cfast and Cslow first; lower bath level to cut pipette capacitance.
- Increase the compensation lag (e.g. 10–20 µs).
50 Hz hum and drifting baseline
- Star grounding to a single point; check the Faraday cage and remove unused mains devices.
- Rechlorinate Ag/AgCl wires and the bath electrode; use an agar bridge if the bath Cl⁻ changes.
- Keep the perfusion line away from the headstage and avoid dripping.
Equation sheet
Every relation in this guide, with a button that copies the LaTeX source for your thesis or slides.
Constants: R = 8.314 J mol⁻¹ K⁻¹, F = 96 485 C mol⁻¹, kB = 1.381 × 10⁻²³ J K⁻¹, e = 1.602 × 10⁻¹⁹ C. RT/F = 25.4 mV at 22 °C, 26.7 mV at 37 °C.
Where these equations come from
- Neher E, Sakmann B (1976). Single-channel currents recorded from membrane of denervated frog muscle fibres. Nature 260:799–802. doi:10.1038/260799a0
- Hamill OP, Marty A, Neher E, Sakmann B, Sigworth FJ (1981). Improved patch-clamp techniques for high-resolution current recording from cells and cell-free membrane patches. Pflügers Arch 391:85–100. doi:10.1007/BF00656997
- Hodgkin AL, Huxley AF (1952). A quantitative description of membrane current and its application to conduction and excitation in nerve. J Physiol 117:500–544. doi:10.1113/jphysiol.1952.sp004764
- Goldman DE (1943). Potential, impedance, and rectification in membranes. J Gen Physiol 27:37–60. doi:10.1085/jgp.27.1.37
- Hodgkin AL, Katz B (1949). The effect of sodium ions on the electrical activity of the giant axon of the squid. J Physiol 108:37–77. doi:10.1113/jphysiol.1949.sp004310
- Sigworth FJ (1980). The variance of sodium current fluctuations at the node of Ranvier. J Physiol 307:97–129.
- Neher E (1992). Correction for liquid junction potentials in patch clamp experiments. Methods Enzymol 207:123–131.
- Barry PH, Lynch JW (1991). Liquid junction potentials and small cell effects in patch-clamp analysis. J Membr Biol 121:101–117.
- Armstrong CM, Gilly WF (1992). Access resistance and space clamp problems associated with whole-cell patch clamping. Methods Enzymol 207:100–122.
- Sakmann B, Neher E, eds. (1995). Single-Channel Recording, 2nd ed. Springer. The standard reference for noise, filtering and dwell-time analysis.
- Molecular Devices. The Axon Guide: Electrophysiology and Biophysics Laboratory Techniques. Free amplifier-level treatment of compensation, noise and grounding.